o1.
Time constant of output filter
Default setting: 0 s (inactive).
To specify a time constant: Select
with key, edit number with
key, proceed with enter
Time constant of output filter (attenuation)
To smoothen the current output, a low-pass filter with
adjustable filter time constant can be switched on. When there
is a jump at the input (100 %), the output level is at 63 % after
the time constant has been reached.
The time constant can be set from 0 to 120 sec.
If the time constant is set to 0 s, the current output follows the
input.
Note:
The filter only acts on the current output, not on the display!
Time constant 0 to 120 sec